Autorzy: Mahmoud Nasr, Jan K. Argasiński, Krzysztof Brzostowski, Adam Piórkowski
Opublikowano w: Computational Collective Intelligence (ICCCI 2026), Lecture Notes in Computer Science (LNAI), vol. 16909, Springer, pp. 316–329. DOI: 10.1007/978-3-032-36871-3_22
Link: https://link.springer.com/chapter/10.1007/978-3-032-36871-3_22
Abstrakt:
Choosing a reconstruction kernel in computed tomography always involves a compromise: sharper kernels reveal finer anatomical detail but amplify noise, while softer kernels yield cleaner images at the price of blurred edges. This chapter introduces BEMD–QBF, a hybrid post-processing method that transforms images reconstructed with a sharp kernel so that they resemble soft-kernel reconstructions, while keeping clinically relevant structures intact. Each image is first split into intrinsic mode functions using Bidimensional Empirical Mode Decomposition, and the resulting components are then smoothed with a Quaternion Bilateral Filter operating in a single quaternion representation, which makes it possible to suppress noise without sacrificing structural detail. The method was tested on images reconstructed with six kernels of varying sharpness (B50, B46, B41, B36, B35 and B31) and benchmarked against Non-Local Means, Anisotropic Diffusion, classical Bilateral Filtering and the standalone Quaternion Bilateral Filter. Measured with SSIM and PSNR, BEMD–QBF achieved the highest structural fidelity for every kernel considered, while its denoising performance remained on par with the competing filters. Since it operates directly on already reconstructed images, the approach offers a practical way to improve CT image quality in settings where raw projection data are not available.